12
Guidance and manufacturer’s declaration – electromagnetic immunity
Össur Myoelectric Prosthetic Devices are intended for use in the electromagnetic environment specified below.
The customers or the users of Össur Myoelectric Prosthetic Devices should assure that each are used in such an
environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Not applicable
Not applicable
Battery Powered
No Cables
>3m
Portable and mobile RF communications
equipment should be used no closer to any part of
the Össur Myoelectric Prosthetic Devices including
cables, than the recommended separation distance
calculated from the equation appropriate to the
frequency of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.7 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site survey
a
should be less than the compliance level in each
frequency range
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF IEC
61000-4-3
12 V/m
26 MHz to
1000 MHz
10 V/m
1000 MHz to
2700 MHz
12 V/m
26 MHz to
1000 MHz
10 V/m
1000 MHz to
2700 MHz
1kHz 80% AM
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Össur Myoelectric
Prosthetic Devices are used exceeds the applicable RF compliance level above, Össur Myoelectric Prosthetic
Devices should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re- orienting or relocating the Össur
Myoelectric Prosthetic Devices
Содержание i-Digits Access
Страница 1: ...i DIGITS Instructions for Use...
Страница 3: ...3 1a 2 3 1b 5 4 1 2 3 4 5...
Страница 108: ...108 i Digits My i Limb Biosim Apple Store Apple iOS iPhone iPad Apple Store i Digits i Digits i Digits i Digits 2...
Страница 109: ...109 i Digits i Digits ssur ssur ssur i Digits ssur ssur ssur ssur 15 600 800 mAh 3 LED LED 90 3...
Страница 114: ...114 ssur ssur CISPR 11 ssur RF CISPR 11 ssur IEC 61000 3 2 IEC 61000 3 3...
Страница 118: ...118 BF IP22 12 5 mm i Digits i Digits 2 6 II...
Страница 179: ...179 i Digits My i Limb Biosim Apple Store Apple iOS iPhone iPad Apple Store i Digits i Digits i Digits i Digits 2...
Страница 180: ...180 i Digits i Digits ssur ssur ssur i Digits ssur ssur ssur ssur 15 600 800 3 90 3...
Страница 185: ...185 ssur Myoelectric ssur Myoelectric CISPR 11 ssur Myoelectric CISPR 11 B ssur Myoelectric IEC 61000 3 2 IEC 61000 3 3...
Страница 189: ...189 BF IP22 12 5 i Digits i Digits 2 6 II...
Страница 192: ...192 i Digits ssur ssur ssur ssur 15 1 800mAh 1 600 1 3 LED LED LED 90 3 4 2 5 2 5 1 4 LED LED...
Страница 199: ...199 BF 4 IP22 12 5 mm i Digits i Digits 2 6 II...
Страница 208: ...208 ssur...
Страница 209: ...209 BF IP22 12 5 mm i Digits i Digits 2 6 II...
Страница 212: ...212 i Digits ssur Vaseline ssur ssur ssur 15 800mAh 600 3 LED LED LED 90 3 4 2 5 2 5 1 4 LED...
Страница 220: ...220 BF YYYY IP22 12 5mm i Digits i Digits 2 6 Class II...
Страница 222: ...222 i Digits i Digits i Digits i Digits 2 i Digits i Digits ssur ssur ssur i Digits ssur...
Страница 230: ...230 BF YYYY 12 5 IP22 i Digits 6 i Digits 2...