11
6.2 Source Measure Units (SMU 1 and SMU 2)
The source measure units output a voltage and then measure both the voltage and current. The output
voltage is always measured at the output of the BNC (rather than assuming it is at the set voltage) in case of
any load effects (for example short circuiting the output or low impedances causing a small drop in voltage).
Each source measure unit has manually selectable ranges so that both high and low currents can be
measured accurately.
6.2.1 Power Specifications
The unit is specified for a maximum operation power as in
Table 6.1
.
Figure 6.1
shows the four-quadrant
safe operation zone of the unit.
Table 6.1
. DC power specification.
Figure 6.1
. The four-quadrant safe operation zone.
Power per SMU channel (continuous)
Four-quadrant Operation Source / Sink
2W (± 10 V at ±200 mA)