13
Table 6.3.
Voltage measurement specifications.
Mode
OSR
Range
Accuracy
Noise
Resolution
Sourcing
N / A
± 10 V
± 10 mV
333 μV
1.7E-4
Measuring
1 to 4
± 10 V
± 10 mV
100 μV
1.0E-4
5 to 9
± 10 V
± 10 mV
50 μV
1.0E-5
1
OSR 10-19 will set the ADC in 2x mode with the same parameters of OSR 0-9, this will approximately double
the unit’s measurement rate.
2
ADC noise increases by approximately √2 when OSR is decreased. Note that this is not the unit
measurement noise levels.
3
Values based on 1000 data points per measurement.
6.2.3 Voltage Source and Measurement Specifications
Warning:
Do
NOT
attach external voltage sources to either SMU module as this may result in
damage to the unit.
6.2.4 Current Measurement Specifications
The current measurement of the Ossila Source Measure Unit is spread over 5 ranges. Each range has
a different maximum current, accuracy, and resolution. As the range number increases, the maximum
measurable current decreases, but the accuracy and resolution increase.
Table 6.4.
Current measurement specifications.
Range
Max current
Accuracy
1
Noise
2
Resolution
Burden
1
± 200 mA
± 500 μA
10 μA
1 μA
< 10mV
2
± 20 mA
± 10 μA
1 μA
100 nA
< 10mV
3
± 2 mA
± 1 μA
100 μA
10 nA
< 10mV
4
± 200 μA
± 100 nA
10 nA
1 nA
< 10mV
5
± 20 μA
± 10 nA
1 nA
100 pA
< 10mV
1
Accuracy has been measured at the maximum current of that range.
2
Noise measurement at the highest OSR (9).