17. MAINTENSNCE
1) Cleaning the exterior
2) Changing the fan filter
3) Cleaning the cartridge compartment
1) Disconnect the analyzer before cleaning.
2) Dust the Analyzer by wiping with a soft, clean cloth.
3) Do not use liquid or aerosol cleaners.
4) If necessary, alcohol swab can be used to wipe out dirt, but other chemical
solutions should not be used. Cleaning solutions may cause damage to the
analyzer surface.
STEP 1. Turn the power off and disconnect the power cord before cleaning.
STEP 2. When the lid is opened, press the
two black knobs to rotate the cartridge
compartment forward.
STEP 3. Dampen a sponge swap with water or ethanol.
STEP 4. Gently wipe the cartridge compartment. Be careful not to drip any
liquids into the system.
Cleaning the cartridge compartment
Cleaning the analyzer
Recommended maintenance routine
17. MAINTENSNCE
Filter
Filter Cover
NOTE
Additional Fan Filter can be purchased from our representatives, nearest in
your area.
Check the fan filters located on the backside of the analyzer regularly. Replace the
fan filter if it is clogged by dust. Extra filters come with your system.
STEP 1. Remove the fan filter cover.
STEP 2. Remove the used fan filter and replace
with a new one carefully.
STEP 3. Close the fan filter cover.
Replacing the Fan Fileter :