11. TESTING HbA1c
STEP 4.
Gently insert the cartridge into the cartridge compartment when ‘Insert Test Cartridge’ is
shown. Hold the cartridge with barcode facing left.
Ensure a gentle snap is either heard or felt to confirm proper placement.
STEP 5.
The display will show the ‘'Apply sample to sample area'' and 'Insert Reagent Pack' icon.
NOTE :
Do not force the cartridge into the analyzer. The cartridge is
designed to fit only in one way.
13/03/30
AM10:00
13/03/30
AM10:02
13/03/30
AM10:02
Gently mix the reagent pack 5 ~ 6 times before applying blood sample.
STEP 5-1.
Apply the blood sample by gently touching the drop of
blood with the tip of the sampling area.
Ensure that the sampling area is completely filled.
STEP 5-2.
11. TESTING HbA1c
CAUTION :
Do not mix too vigorously, it may cause air bubbles. If bubbles are
present wait until they disappear before testing.
IMPORTANT :
Once the reagent pack is filled with the blood sample, analysis must
begin immediately.