22
23
11. TESTING HbA1c
Insert the reagent pack into the cartridge in the analyzer. The ‘close the lid’ icon will be
displayed.
STEP 6.
The test starts automatically when the lid is closed.
STEP 7.
NOTE :
Do not force the reagent pack into the cartridge, it will only fit one way.
IMPORTANT :
Do not open the lid during test.
Do not vibrate or move the analyzer during test.
AM10:03
AM10:03
13/03/30
4:30
13/03/30
11. TESTING HbA1c
The measuring time is 5 minutes and the test result will be displayed (in % or mmol/mol).
STEP 8.
After the test is completed, open the analyzer lid. ‘Remove cartridge’ will be shown.
Remove the test cartridge from the analyzer by gently pushing it to the left and
pulling it out.
STEP 9.
CAUTION :
Do not force the cartridge to remove it from the analyzer. Dispose all
waste in accordance with applicable national and/or local regulations.
NOTE :
If the results “>14%” or “<4%” are displayed, repeat testing to confirm
result. If the second result also is outside the range, contact your local
distributor.
ID:
AM10:09
13/03/30
5.4
%
13/03/30
Remove
AM10:08