ATTENTION WARNING:
Complying with the following in-
structions will help reduce unnec-
essary wear and damage to the
ARTROMOT
®
-F.
1. ALWAYS use the hand-held pro-
gramming unit to run the device
to adjust the position of the Drive
Bar and Finger Actuators.
2. DO NOT force the actuators to
move by hand by pushing on the
Drive Bar. This action breaks the
gear in transmission and requi-
res the gear motor to be repla-
ced. This action also breaks the
Arm Drive Gear Support Pins
and requires that each damaged
Bridge Arm assembly be replaced.
(fig. 23/24)
Electrical connection:
Battery 6V 4AA
100-240 V,
50/60 Hz
Rated current motor:
2.5 A max.
Power consumption:
40 VA
Fuses:
1.5 A PTC
Degree of protection:
VDE
Protection
class I,
Type B
Reverse-on-load:
1.6 kg low
4.3 kg high
Weight:
Hand-held
programming
unit: 0.2 kg
Drive assembly
with patient kit:
0.66 kg
MPG:
Class 2a
CSA:
IEC 60601-1
CSA 601.1-M90
UL 2601-1
Ambient conditions
Ambient temperature
-40°C to +70°C
Relative air humidity
10% to 100%
Air pressure
500 hPa to
1060 hPa
Operational conditions
Ambient temperature
+10°C to +40°C
Relative air humidity
30% to 75%
Air pressure
700 hPa to
1060 hPa
Subject to change without notice (12/03).
26
English
8. Specifications
Содержание Artromot-F
Страница 1: ...ARTROMOT F Gebrauchsanweisung Operating Instructions...
Страница 2: ...Ger tebeschreibung 2 Device description 6 10 8 7 9 1 3 4 5 20 18 17 16 11 2 19 12 13 15 13 14...
Страница 4: ...32 Abbildungen Figures 1 2 3 4 5 6 7 8...
Страница 5: ...33 9 10 11 12 13 14 16 15...
Страница 33: ...Notes 31...