Moxi Flow
™
User Guide
Page 15
Re-scaling the x axis (size range)
Upon completion of a test (but prior to saving the test), the x-axis scale can be adjusted. This is
accomplished by selecting the re-scale button below the scatter plot (magnifying glass with
arrow, see image sequence for an MF-S cassette below). With each button press, the text
under the button changes to indicate the scale range that will be displayed on the next button
push. The possible size scale ranges are:
•
MF-S Cassettes:
2-25
µ
m
,
2-17
µ
m
and
2-9
µ
m
.
•
MF-M Cassettes:
3-33
µ
m
,
3-25
µ
m
,
3-17
µ
m
, and
3-9
µ
m
Rescaling the axis results in reprocessing of the raw data to ensure that resolution is not lost
during the rescale process (1000 histogram bins). This allows (and is recommended for) users
to achieve finer resolution of smaller particle populations and potentially increased separation
between particle sub-populations.
However, once a test has been saved, rescaling the x-
axis is not a possibility.
Removing the noise region
The Moxi Flow automatically identifies the noise region and removes it. To manually adjust the
noise gate, press the red gate button at the top right of the screen and toggle until the yellow
noise gate appears (see left image below). The yellow noise marker can then be dragged
around the noise region.
Once the test has been saved, the noise region is permanently
removed and the noise gate cannot be re-adjusted.
Touch Rescale
Touch Rescale
Toggle
Noise Gate
Size Gates
Fl. Gate