- 013
2.
Operation Procedure
2.1.
Start up Procedures
❖
Turn on Mechanical pump and Air compressor (make sure V1 is closed). The pressure of
compressor should over 4kgf/c
m
2
(0.4MPa).
❖
Turn on Main switch and Main frame switch.
❖
Switch
“AUTO/MAN”
to
“AUTO”
at the vacuum control panel which has been installed in the
rear part of Main frame.
❖
Turn on Computer, run the software of SEM.
❖
The password of windows operation system is
“sem”.
❖
Click the [Start] button, the color of Start button will turn red, the vacuum system will come
into the working status (V2 will be turned on)
❖
After the indicator lamp of V2 V3 TMP went on, vacuum system come into the high vacuum
working status. About 5 mins the vacuum indication turns green, the system can begin to
work. At this moment, the display of vacuum will reach over 3.0*10
-5
Torr.