OPTILAB EYDFA-33-PM-R Скачать руководство пользователя страница 10

EYDFA-33-PM-R Operation Manual  

 

R1.00 

 

 

Page 7 of 15 

 

 

Optilab, LLC

 

 

600 E. Camelback Road, Phoenix, AZ  85012 
Phone:  1-877-303-3888, Fax:  (602) 343-1489, Email:  [email protected] 

 

 
 

 

 

Feature 

Function 

 

 AC Power Socket

 

 
The AC power socket is the input for the AC power source.  
A three-pin standard power cord should be used to connect 
this equipment to any 110 or 220 V main supply.   
 

 Main AC Power 

Switch

 

This switch enables the electrical power to the DFB-B unit.

 

 
 

 

USB Female Socket  Using a  USB  cable, this  port allows for  remote  control  and 

monitoring through a PC workstation. 
 

 Ventilation Fans 

The ventilation fans ensure proper ventilation inside the 
unit.  The back panel and the two sides of this equipment 
should be placed at least 3 inches from the wall to dissipate 
heat effectively.  
 

 Interlock 

This BNC female connector is a safety interlock. It must be 
shorted in order for the pump lasers to enable. Use the 
provided accessory to short this port or connect it to a 
compatible interlock device. If the interlock is open during 
normal operation, the pump lasers will be turned off. To re-
enable the pump laser output, turn the output switch on the 
front panel to “OFF” position first and start over.

 

 

 
 
 
 
 
 

Rear Panel 

Содержание EYDFA-33-PM-R

Страница 1: ...THER THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE THE USER MUST READ THIS MANUAL BEFORE OPERATING THE PRODUCT OPERATIONS OTHER THAN THOSE DESCRIBED IN THIS MANUAL MAY RESULT...

Страница 2: ...00 08 08 2018 Manual introduced Optilab LLC 2018 All rights reserved Information in this document is subject to change without notice All pictures and diagrams are for illustration purposes only and t...

Страница 3: ...__________________6 3 2 Operation Instructions ______________________________________________________8 3 2 1Preparation Procedure ___________________________________________________8 3 2 2Start up Pro...

Страница 4: ...report There MUST be an input signal present before turning on the optical output Do NOT use this product in HIGH PEAK PULSE POWER 10W applications unless with the consent from the manufacturer Faili...

Страница 5: ...AL COMPONENTS CONNECTED TO THE OPTICAL OUTPUT MUST BE RATED FOR HIGH POWER USE It is highly recommended that these components be properly cooled to dissipate heat FOR CONTINUED PROTECTION AGAINST FIRE...

Страница 6: ...erformance Standards for Light Emitting Products 5 US Department of Labor OSHA Publication 8 1 7 Guidelines for Laser Safety and Hazard Assessment 6 Laser Safety Equipment Laurin Publishing 7 Laser sa...

Страница 7: ...and up to 33 dBm saturation power at 1550 nm It is a self contained turn key rackmount designed to supply high and stable gain across the specified range of the wavelength window 2 2 Features Up to 33...

Страница 8: ...put Power Stability 0 2 dB over 8 hours Optical Input Output Fiber Type 8 125 NA 0 12 PM Optical Input Output Connector FC APC fast axis aligned to slow axis Mechanical Specifications Operating Temper...

Страница 9: ...The Touch LCD display shows the parameters and optical settings of the EYDFA rackmount unit This is a touchscreen type display see section 3 2 3 for more information Output Power Enable Switch This sw...

Страница 10: ...remote control and monitoring through a PC workstation Ventilation Fans The ventilation fans ensure proper ventilation inside the unit The back panel and the two sides of this equipment should be plac...

Страница 11: ...off and make sure the key switch on the front panel is also off Interlock should be shorted using the provided accessory or connected to a compatible interlock device After plugging in the appropriat...

Страница 12: ...ight a parameter to change it will turn yellow When the parameter is highlighted use the INC button to cycle through the options For the pump current adjust each digit then press SAVE to engage the ne...

Страница 13: ...ry to short this port or connect it to a compatible interlock device If the interlock is open during normal operation the pump lasers will be turned off To re enable the pump laser output turn the out...

Страница 14: ...the command afterwards Syntax is CMD XX case sensitive READ This command displays system status including the current settings of each pump and maximum current setting Example of Response Optilab LLC...

Страница 15: ...input or optical input power is too small S Check whether optical input is present or check optical input power is correct according to the test report C Optical connectors are dirty S Disable optica...

Страница 16: ...moist wipe removes dust particles oil and contaminants that may damage or blot the end face of the connector during connection The dry wipe removes residual alcohol that may be ignited by optical emis...

Страница 17: ...posure or damage to the optical connector by optical emission 2 Moisten the swab by placing it on top of the alcohol dispenser and push down to saturate the swab 3 For receptacles adapters or other co...

Страница 18: ...effect all repairs under terms of the warranty providing the Product is returned to Optilab LLC as specified in the warranty statement The purchaser shall bear the freight charges incurred in returnin...