Cleaning
30
11
Cleaning
Set the sensor cleaning setting from SC-U1.
To perform cleaning, press
.
Manual cleaning
Activate the wiper manually. While cleaning the measurement value of the
target sensor is retained (held).
Cleaning interval
Set a time interval to activate the wiper.
This product is a sensor with wipers and it operates as following.
・
The setting time for the cleaning interval is the time interval between the
end of cleaning and the next cleaning.
・
The cleaning action is fixed at 1 wiper blade round trip per cycle.
Item
Setting range
Factory default
setting
Cleaning Interval OFF (No cleaning)
10 to 1440 minutes (by 10 minutes)
30 minutes
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "9 Cleaning" and "
◆
Manual Cleaning".
Select an item from the following.
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "9 Cleaning" and "
◆
Cleaning Interval".
Caution
・
Pressing
automatically retains the measurement value.
・
Do not press
while a sensor with a wiper is in the air.
This may cause the sensor to malfunction.
Cleaning
1 round
trip of
wiper blade
1 round
trip of
wiper blade
Cleaning
Cleaning not performed
Cleaning not performed
Cleaning interval
Cleaning interval