Relay Setting
17
7
Relay Setting
Set the relay output setting from SC-U1.
The factory default setting is OFF for relay output.
To set the relay setting, press
.
Set the alarm to be output when the measurement value exceeds a certain
value.
The following settings are available.
Alarm output setting
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.4 Relay Setting" and "
◆
Alarm output setting".
Setting item
Setting range
Suspended solids [SS]
Transparency [Trans.]
Trigger
High Alarm/Low Alarm
Active Level
0.0 to 1000
0 to 100
Hysteresis High Alarm 0.0 to Active level
0 to Active level
Low Alarm 0.0 to (1000 - Active
level)
0 to (100 - Active level)
ON Delay
OFF
1 to 120 minutes
OFF Delay
OFF
1 to 120 minutes
The setting unit for the active level is based on the display resolution of the item
set for the display item. (For display resolution, refer to “14 Specifications” P.34)
Caution
Pressing
automatically retains the measurement value.
Caution
When “Display item” P.13 is changed, the output item of the
channel of relay setting that has set the display item becomes OFF.
Set again if necessary.