Sensor Setting
14
Response time
Set a signal output response time.
A measurement value is calculated using moving average of the time period
set by the signal output response time.
For example, if the sensor measures values every 5 seconds and the signal
output response time is set to 50 seconds, which is 5 seconds
×
10 times,
the moving average of 10 times measurement is calculated and used as the
measurement value.
The recommended value for this product is 30 seconds or longer.
If the response is too fast with the factory default setting (5 seconds),
change the signal output response time as needed.
Setting range
Factory default
setting
Response
Time
5 to 600 seconds (by 5 seconds)
5 seconds
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.1 Sensor Setting" and "
◆
Response Time".
The following display resolutions are available.
SS
・
0.0 - 49.9: 0.1
・
50 - 1000: 1
Transparency
・
0 - 100: 1
Measurement item
Display item
Factory default
setting
Item
Unit
Main
display
Suspended solids
SS
mg/L
SS mg/L
SS
No unit
Transparency
Trans.
cm
2nd
display
-
OFF
-
OFF
Suspended solids
SS
No unit
SS
mg/L
Transparency
Trans.
cm