50
4-2-5 Setup Menu (
setup
)
Operating Mode
All parameters will be set to default values if any change in made in this parame-
ter, except to those for the communications and protect settings.
f1
: Displays the reciprocal of the time (T) between the moment input A is
turned ON and the moment input B is turned ON after being multiplied by
60.
f2
: Measures and displays the cycle (T) of input A as is.
f3
: Displays the time between the moment input A is turned ON and the mo-
ment input B is turned ON as is.
f4
: Displays the elapsed time (T) of input A.
f5
: Displays the number of pulses of input A counted while input B is ON.
f6
: Displays the number of pulses of input A counted between the moment in-
put B is turned ON and the moment input B is turned ON again.
Setting
Default
f1
: Passing speed
f2
: Cycle
f3
: Time difference
f4
: Elapsed time
f5
: Length measurement
f6
: Interval
f1
Refer to
Section 3 Operating Modes
.
Follow the steps described below to select the
f4
elapsed time setting.
Set Value LED Display Model
Basic Model
1, 2, 3...
1. Press the Mode Key for more than one second while the
setup
setup menu
is displayed. The
func
operating mode setting will appear.
Set Value LED Display Model
Basic Model
2. Press the Shift Key to display
f1
for changing. The PROG indicator will flash.
Set Value LED Display Model
Basic Model
func
FUNCTION
SETTING
REFERENCE
SETTING
EXAMPLE
Setting Mode
Section 4-2