46
4-2-4 Prescaling Menu (
pscl
)
Bank No. of Prescale
Prescaling Value X (Mantissa) of Input A
Prescaling Value Y (Exponent) of Input A
Decimal Point Position
The measured data of passing speed or cycle is displayed after being multiplied
by a preset value (i.e., prescaling value).
Display value = Measured data x Prescaling value
Input type
Setting
Default
p.bank
: Bank no. of prescale
off
/1 to 4
off
ps
*
.ax
: Prescaling value X (mantissa)
of input A
0.0001 to 9.9999
1.0000
ps
*
.ay
: Prescaling value Y (exponent)
of input A
--9 to 9
0
decp.
*
: Decimal point position
One of the 1st to 4th digits
from the right
No decimal
point
position
setting
Refer to
Section 3 Operating Modes
.
Follow the steps described below to input the following.
Operating mode = F1
Prescaling bank = OFF
Prescaling value X (mantissa) of input A = 0.5000
Prescaling value Y (exponent) of input A = --1
Decimal point =
jjjj
.
j
(First digit from the right)
Set Value LED Display Model
Basic Model
1, 2, 3...
1. Press the Mode Key for more than one second while the
pscl
prescaling
menu is displayed. The
p.bank
prescaling bank setting will be displayed.
Set Value LED Display Model
Basic Model
p.bank
ps
*
.ax
ps
*
.ay
decp.
*
FUNCTION
SETTING
REFERENCE
SETTING
EXAMPLE
Setting Mode
Section 4-2