7. Press the Shift Key to select whether or not to accept the settings.
Set Value LED Display Model
Basic Model
8. Press the Up Key to select
yes
.
9. Press the Mode Key to overwrite all calibration data and to put the K3NH into
RUN mode. The calibration data will not be overwritten if
no
is selected.
Set Value LED Display Model
Basic Model
RUN mode
Calibration (0 to 10 V)
1, 2, 3...
1. First, the
sure
alarm display will appear. This will not appear once field cal-
ibration is performed.
Set Value LED Display Model
Basic Model
2. Press the Mode Key to display
2U 10
. The teaching indicator will be lit.
Set Value LED Display Model
Basic Model
3. Press the RESET/TEACH Key to display the calibration value setting. Then
set the output of STV to 10 V. The teaching indicator will flash. After the value
stabilizes (i.e., the fluctuation of the value is less than several digits), press
the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
4. The display will change to
2U 0
for 0-V calibration. The teaching indicator will
be lit.
Set Value LED Display Model
Basic Model
Maintenance Mode
Section 5-3