82
value stabilizes (i.e., the fluctuation of the value is less than several digits),
press the Mode Key to tentatively register the calibration data.
Set Value LED Display Model
Basic Model
10. Calibrate the bias compensation value as the last item. The display will
change to
bias
. The teaching indicator will be lit. Disconnect the STV, set
the cold junction compensator to 0
°
C, and enable the built-in thermocouple.
Be sure to disconnect the wires connected to the STV.
Set Value LED Display Model
Basic Model
11. Press the RESET/TEACH Key to display the calibration value setting. The
teaching indicator will flash. After the value stabilizes (i.e., the fluctuation of
the value is less than several digits), press the Mode Key to tentatively regis-
ter the calibration data.
Set Value LED Display Model
Basic Model
12. The display will change to
str
for confirmation of the settings.
Set Value LED Display Model
Basic Model
13. Press the Shift Key to select whether or not to accept the settings.
Set Value LED Display Model
Basic Model
14. Press the Up Key to select
yes
.
15. Press the Mode Key to overwrite all calibration data and to put the K3NH into
RUN mode. The calibration data will not be overwritten if
no
is selected.
Set Value LED Display Model
Basic Model
RUN mode
Maintenance Mode
Section 5-3