SECTION 5. MOTOR UNIT MAINTENANCE
5.1 MOTOR CLEANING
The MOTOR housing may be wiped down with a clean cloth when
disconnected from the power source and not in use. DO NOT use
solvents to clean the unit.
5.2 BRUSH REPLACEMENT
1. Remove Power.
2. Using the driver end of the collet wrench, unscrew and remove
one BRUSH CAP.
3. Pull on the brush spring. The BRUSH and spring will come out as one
assembly. If the spring breaks, turn the MOTOR on its side and tap it
until the BRUSH falls out.
4. Repeat steps 2 and 3 for the other BRUSH.
5. Install the new BRUSHES (with springs) into the MOTOR unit.
6. Re-install the BRUSH CAPS to both sides of the MOTOR unit and
tighten with the collet wrench. The Tissue Master is now ready for use.
WARNING: DO NOT
connect the power cord to a power source if it is
worn or damaged. For assistance, call your authorized dealer or Omni
International at 1-800-776-4431 of 770-421-0058.
WARNING: DO NOT
attempt to disassemble the motor unit.
11
BRUSH (2)
BRUSH CAP (2)