1.3 STAND-MOUNTING THE TISSUE MASTER (Optional)
An optional stand kit (Part # F22000) may be purchased to mount the
Tissue Master. Contact your authorized dealer or Omni International
at 1-800-776-4431 or 1-770-421-0058 to order.
Assemble the stand:
1. Thread the larger rod onto the base plate.
2. Thread the cross-rod into the motor cradle.
3. Assemble the clamp to the vertical rod.
4. Assemble the cross-rod to the clamp.
5. Insert the motor unit into the motor cradle
and tighten the thumb screws.
1. Check that the power switch is in the OFF position.
2. Plug the power cord into an appropriate outlet.
WARNING!: DO NOT
attempt to plug a cord into an incompatible
outlet.
3. The bottom of the GENERATOR must be submersed in a liquid
before operation to prevent bearing damage.
4. Turn ON the power switch. The speed progressively increases as the
switch slides away from the “O” mark. Run the motor for the
appropriate amount of time required to homogenize the sample.
Moving the homogenizer up and down within the sample during
processing will increase processing efficiency.
5. A simple cleaning of the GENERATOR in between samples may be
accomplished by running the probe in a chlorine bleach solution and
then in distilled water to rinse. The probe must still be cleaned
periodically in accordance with section 3.
WARNING: DO NOT
process pathogenic material in an open
container. For assistance, contact your authorized dealer or
Omni International at 1-800-776-4431
SECTION 1. PREPARATION FOR USE
SECTION 2. OPERATION
6
Содержание TISSUE MASTER 125
Страница 1: ...OMNI TISSUE MASTER USER MANUAL REV B ...
Страница 14: ...NOTES 14 ...
Страница 15: ...NOTES 15 ...