9
Safety Instructions
2.3 Operating the CP CR500
For safe
CP CR500
operation:
1. Make sure that only the green light "0" on the
CPC 100
's front panel is lit.
CP TD1
may not give any acoustic warning sound.
2. Secure the
CP CR500
unit with a chain or a barrier tape at a distance of at
least 1 m / 3 ft to prevent an accidental entering of the dangerous zone
during operation.
3. Connect
CP CR500
,
CPC 100
and
CP TD1
as described in this user manual.
4. Do not operate the
CP CR500
without connecting the safety circuit!
5. Make sure that the connectors are tighlty fit before turning on the high
voltage.
6. Turn on the high voltage only for the short time that the measurement
requires.
7. Do not touch cables while operating.
8. Do not leave the test set during operation.
9. Once the measurement is finished, press the emergency stop button on the
CPC 100
's front panel.
10.After you have finished working with the
CP CR500
, turn the safety key on
the
CPC 100
's front panel to "lock" (vertical) and remove the key to avoid
anyone turning on the high voltage accidentally.