
CP CR500
8
– Properly ground
CP CR500
,
CP TD1
and
CPC 100
by connecting them to
the common equipotential ground. Use earth connection cables of at least
6 mm² cross-section. Never operate the
CP CR500
without solid grounding.
– Never touch the high-voltage sockets of the
CP CR500
during operation.
Consider these sockets as live, and due to the high voltage life-hazardous.
– The plugs of the high-voltage cables and the cables themselves are not
shielded and therefore not safe. While operating the
CP CR500
, consider
these cables as live, and due to the high voltage life-hazardous.
2.2 Handling the High-Voltage Cables and the
CP CR500
Before handling the high-voltage cables or the
CP CR500
proceed as follows:
1. Press the emergency stop button on the
CPC 100
's front panel or disconnect
CPC 100
from the mains.
2. While working in the dangerous zone, turn the safety key on the
CPC 100
's
front panel to "lock" (vertical) and remove the key to prevent anyone from
accidentally turning on the high voltage.
3. Check the operational condition, in particular:
– that only the green light "0" on the
CPC 100
's front panel is lit and
–
that CP TD1
does not give an acoustic warning sound.
4. Now the high-voltage cables and the
CP CR500
can be handled safely.
12 kV
Содержание CP CR500
Страница 6: ...CP CR500 6...
Страница 10: ...CP CR500 10...
Страница 18: ...CP CR500 18...
Страница 38: ...CP CR500 38...
Страница 40: ...CP CR500 40...
Страница 42: ...OMICRON Contact Addresses 42...