12
3.3 Low-Voltage Connector (LV)
The
CP CR500
provides one low-voltage connector LV for connecting with
CP TD1
and, if necessary, to another
CP CR50
3.4 Safety
Connection
On the interface panel of the
CP CR500
, there are two D-Sub 9 connectors
(male and female) named Safety A (male) and Safety B (female).
These connectors are used for the safety circuit.
3.5 Equipotential
Ground
The connector EG is connected with the
CP CR500
's equipotential ground
(functional earth, protective earth).
Caution:
For a safe operation make sure that all units involved and the trolley, if used, are
connected to the common equipotential ground. Use earth connection cables of
at least 6 mm² cross-section. Never operate the test setup without solid
grounding.
Figure 3-1:
CP CR500
’s Interface
Panel
Safety B
(D-Sub 9
female)
Safety A
(D-Sub 9 male)
LV connector
(low voltage)
Connector for
equipotential
ground