8.3.4 Calibrating a Shunt-Thru or Series-Thru measurement
In this section you learn how to calibrate an Impedance, Reflection or Admittance measurement in the
Shunt-Thru
or
Series-Thru measurement mode
.
Shunt-Thru and Series-Thru are based on a S21 Transmission measurement.
Bode 100
measures
S21 and the
Bode Analyzer Suite
calculates impedance from the S21 measurement. Details on the
You can either calibrate the underlying S21 measurement using Thru calibration
or
you can use Open/
Short/Load calibration to move the reference plane directly to the DUT.
Shunt-Thru and Series-Thru allow
both
, Thru
or
Open/Short/Load calibration.
However, only one calibration can be active at a time!
Even if both calibrations have been performed only one calibration is applied.
You must select the calibration you want to apply using the slider in the calibration dialog.
If the arrow points to the left, Thru calibration is applied.
If the arrow points to the right, Open/Short/Load calibration is applied (see example below).
Shunt-Thru calibration connections
Calibrating Thru:
Thru calibration can e.g. remove the gain and
phase error introduced by the connection cables
including a coaxial common mode transformer
that is generally used in this measurement to
suppress the cable-braid error.
Calibrating Open/Short/Load:
Open calibration.
84
OMICRON Lab
Содержание Bode 100
Страница 1: ...Smart Measurement Solutions Bode 100 User Manual...
Страница 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Страница 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Страница 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...