Connect Open and press
Start
.
Wait until the Open calibration has performed.
Check if your Short element fits to the Short
Delay Time Setting (50 ps default).
Connect Short and press
Start
.
Wait until the Short calibration has performed.
Check if your Load element fits the Load
Resistance Setting (50 Ω default).
Connect Load and press
Start
.
Wait until the Load calibration has performed.
Calibration
OMICRON Lab
81