Typical applications
Typical applications for the Transmission / Reflection measurement mode are:
•
Measuring Transmission and Reflection of filters such as IF filters or EMI filters.
•
Comparing results from measurements with internal and external reference connection
(the Transmission / Reflection measurement mode is the only measurement mode that allows
changing the Receiver 1 connection).
•
Measuring Gain of RF amplifiers in the 50 Ω domain.
Transmission / Reflection measurement example
Follow the steps described below to perform a Transmission / Reflection measurement:
Connect the test object "IF Filter" to
Bode 100
using two BNC cables as shown in the figure below.
Figure 7-1: Connecting the test object IF Filter to
Bode 100
42