27 |
P a g e
XW-ED and XW-ED-PRO Inputs
Digital Inputs
Type
Range
Accuracy
Frequency
0.01Hz to 100Hz
±0.5%
Frequency
100Hz to 1000Hz
±1Hz averaged over 1s
Counter
0 to +8388608
±1 count max
Up / Down Counter -8388608 to +8388608
±1 count max
Pulse Width (T
pw
)
0.2ms min
±50uS ±1%
Pulse Delay (T
pp
)
0.2ms min
±50uS ±1%
Duty Cycle
1% to 99%
±50uS *F
Process Inputs
Type
Range
Accuracy
Current
0-24mA
±0.1mA
Voltage
0-1V
±5.0mV
XW-EDA and XW-EDA-PRO Inputs
Thermocouple
Type
Range
Accuracy*
J
-210°C to 1200°C
0.4°C
K
-160°C to 1372°C
0.4°C
T
-190°C to 400°C
0.4°C
E
-220°C to 1000°C
0.4°C
N
-100°C to 1300°C
0.4°C
R
40°C to 1788°C
0.5°C
S
100°C to 1768°C
0.5°C
B
640°C to 1820°C
0.5°C
C
0°C to 2320°C
0.4°C
*At 25°C CJC
Accuracy: ±0.05°C per C
RTD
Type
Range
Accuracy
385, 4 Wire
-200°C to 850°C
0.3°C
385, 3 Wire
-200°C to 850°C
0.3°C
385, 2 Wire
-200°C to 850°C
0.6°C
392, 4 Wire
-200°C to 660°C
0.3°C
392, 3 Wire
-200°C to 660°C
0.3°C
392, 2 Wire
-200°C to 660°C
0.6°C
3916, 4 Wire -200°C to 660°C
0.3°C
3916, 3 Wire -200°C to 660°C
0.3°C
3916, 2 Wire -200°C to 660°C
0.6°C
Process Inputs
Type
Range
Accuracy
Current
0-24mA
±10uA
Voltage
0-10V
DC
±5.0mV
Voltage
0-1V
DC
±0.5mV
Voltage
0-0.1V
DC
±0.05mV