Re-Calibration – 5 Point START
To begin the 5-point re-calibration procedure, turn the unit OFF
and follow the steps below.
Keystroke
Display
1.
Press and hold the
MIN/MAX key and turn
the unit on by pressing
the ON/OFF key.
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
2.
Press the
arrow key
until “START” is
displayed on the bottom
line.
Top line reads “RE-CAL”.
Bottom line reads “START”.
3.
Press the PRGM key.
Top line reads “RE-CAL
Bottom line reads “1-
POINT”.
4.
Press the
arrow key
until “5-POINT” is
displayed on the bottom
line.
Top line reads “RE-CAL
START”.
Bottom line reads “5-
POINT”.
5.
Press the PRGM key.
Top line reads “POINT 1 –
ZERO:”
Bottom line displays live
applied pressure.
6.
Vent P1 and P2 ports to
atmosphere and
simultaneously press the
MIN/MAX and HOLD
keys, then release.
Unit takes new zero. Top line
reads “ POINT 1 - ZERO:”
Bottom line displays live
applied pressure. POINT 1
has been taken.
7.
Press the
arrow key
while on the right most
digit to proceed.
Top line reads “ POINT 2 -
APPLY:”.
Bottom line displays the cal
point value to apply.
8.
Apply the indicated
calibration point pressure
using external pressure
standards. After pressure
is stable, press the
arrow key.
Top line reads “ POINT 3 -
APPLY:”.
Bottom line displays the cal
point value to apply.
9.
Repeat step 8 for CAL
POINTS 4 and 5.
Top line reads “POINT 4/5 -
APPLY”
Bottom line displays the cal
point value.