Re-Calibration – 5 Point EDIT
To edit the calibration points for a 5 Point re-calibration follow the
steps below.
NOTE: If the factory default values are acceptable, skip this
section and proceed to the re-calibration 5-Point START procedure.
Keystroke
Display
1.
With unit OFF, press and
hold the MIN/MAX key,
turn the unit on using the
ON/OFF key, then release
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
2.
Press the PRGM key.
Top line reads “CAL POINT
1”. Bottom line displays the
cal point value.
3.
Press the
or
arrow
keys to edit the selected
digit. Use the
or
arrow keys to change the
cursor position.
Note: For 0% go directly
to step 4.
Bottom line displays the cal
point value. The cursor
flashes at the first position
while the value is changed,
then moves to the right
position when the right arrow
key is pressed.
4.
Press the
arrow key
while on the right most
digit to proceed.
Top line reads “CAL POINT
2”. Bottom line displays the
cal point value.
5.
Repeat steps 3 and 4 for
CAL POINTS 2, 3, 4 and
5.
Top line reads “CAL POINT
2/3/4/5”. Bottom line
displays the cal point value.
6.
After editing CAL
POINT 5 press the
arrow key while on the
right most digit to
proceed.
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
7.
To perform the 5-point re-
cal, press the
arrow
key until START is
displayed on the bottom
line.
OR
To exit without
performing the 5-point re-
cal press the
arrow key
Top line reads “RE-CAL”.
Bottom line, “START”.
Continue with 5-Point Re-
calibration procedure at step 3
on next page.
OR
Returns to Measure Mode.