31 of 36
10.
Use
or
arrow keys
to select NO or YES
when asked “Save?” the
Re-Calibration data.
Top line reads “SAVE?”.
Bottom line reads “NO” or
“YES”.
11.
Press the PRGM key at
YES to save the Re-
Calibration data or at NO
to exit without saving.
Top line reads “RE-CAL”.
Bottom line reads “START”.
Re-cal is complete.
12.
Press the
arrow key.
Returns to Measure Mode.
Re-Calibration – Restore Factory Defaults
To restore the re-calibration data to the factory defaults, follow
these steps:
Keystroke
Display
1.
With unit OFF, press and
hold the MIN/MAX key,
turn the unit on using the
ON/OFF key, then
release.
Top line reads “RE-CAL”.
Bottom line reads “EDIT”.
2.
Press the
arrow key
twice.
Top line reads “RE-CAL”.
Bottom line reads “RESTORE
DEFAULTS”.
3.
Press the PRGM key.
Top line reads “RESTORE
DEFAULTS”.
Bottom reads “YES” or “NO”.
4.
Use the
and
arrow
keys to select YES or NO
when asked to restore
defaults.
Top line reads “RESTORE
DEFAULTS”.
Bottom reads “YES” or “NO”.
5.
Press the PRGM key at
YES to restore the
Factory Default
Calibration data or at NO
to exit without restoring.
Top line reads “RE-CAL”.
Bottom line reads “RESTORE
DEFAULTS”. Factory
defaults have been restored.
6.
Press the
arrow key.
Returns to Measure Mode.