Model DFG-RS5 Digital Force / Torque Indicator
User’s Guide
4
Example 1
Model RLC01-100 sensor with Model DFG-RS5 Indicator
RLC01-100
±0.15% of full scale
+
DFG-RS5
±0.1% of full scale
=
Total
±0.25% of full scale
This translates into a fixed error of up to:
0.25% x 100 lbF = 0.25 lbF
Example 2
Model RTQ-50Z sensor with Model DFG-RS3 Indicator
RTQ50-50Z
±0.35% of full scale
+
DFG-RS3
±0.2% of full scale
=
Total
±0.55% of full scale
This translates into a fixed error of up to:
0.55% x 50 ozFin = 0.275 ozFin
Because accuracy is defined as a
percentage of full scale
, the fixed error is possible anywhere on the
scale from 0 to the capacity. As such, this value represents an increasingly large error as
percentage of
reading
towards the low end of the scale. It is, therefore, recommended that a sensor is selected with
capacity as close as possible to the expected load.
The resolution may be different for some sensors depending on which indicator is used. For example, a
Series RLC01 force sensor will display finer resolution when connected to a DFG-RS5 indicator than
when connected to a DFG-RS3 indicator. Resolution information is shown in the sensors’ user’s guide.
1.4 Safety / Proper Usage
Read through the following safety instructions thoroughly before using the DFG-RS5 with a sensor:
1. Note the sensor’s capacity before use and ensure that the capacity is not exceeded. Producing a
load greater than the indicated safe overload value can damage the sensor. An overload can
occur whether the sensor’s indicator is powered on or off.
2. In order to extend the life of the sensor, avoid repetitive shock and impact loading.
3. When moving the sensor to another location, never lift from the cable or strain relief. This can
cause damage to the sensor. Always lift the sensor housing itself.
4. Always ensure that load is applied axially with respect to the sensor.
5. Ensure that the sensor is kept away from water or any other electrically conductive liquids at all
times.
6. The sensor and indicator should be serviced by a trained technician only. AC power must be
disconnected and the indicator must be powered off before the housing is opened.
7. Always consider the characteristics of the sample being tested before initiating a test. A risk
assessment should be carried out beforehand to ensure that all safety measures have been
addressed and implemented.
8. Typical materials able to be tested include many manufactured items, such as springs, electronic
components, fasteners, caps, films, mechanical assemblies, and many others. Items that should
Содержание DFG-RS5
Страница 1: ...Model DFG RS5 Digital Force Torque Indicator...
Страница 2: ...Model DFG RS5 Digital Force Torque Indicator User s Guide 1...
Страница 34: ...Model DFG RS5 Digital Force Torque Indicator User s Guide 33...
Страница 35: ...Model DFG RS5 Digital Force Torque Indicator User s Guide 34 M 5251 1017...