Model DFG-RS5 Digital Force / Torque Indicator
User’s Guide
25
9. The display appears as follows:
Use the
UP
and
DOWN
keys to adjust the load value as required. The load values default to even
increments, as indicated by the previously entered number of data points (even increments are
recommended for best results). For example, if a 50 lbF capacity sensor is calibrated, and 5 data
points were selected, the load values will default to 10, 20, 30, 40, and 50 lb. Apply the calibration
load. Then press
ENTER
.
Repeat the above step for the number of data points selected.
10. After all the compression calibration points have been completed, the display appears as follows:
Press
ENTER
.
11. At the completion of the tension calibration, the display appears as follows:
To save the calibration information, select “Save & exit”. To exit without saving the data select
“Exit without saving”.
12. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
CALIBRATION
Units must be gF.
Please try again
Press ENTER.
CALIBRATION
COMPLETE
Save & exit
Exit w/o saving
CALIBRATION
COMPRESSION COMPLETE
Reverse direction
for tension.
Attach necessary
weight fixtures,
then press ENTER.
CALIBRATION
COMPRESSION
Apply load
1 OF 5
Enter load:
2.000 lbF
Press ENTER.