Model DFG-RS5 Digital Force / Torque Indicator
User’s Guide
23
The sensor can be calibrated at up to 10 points in each direction. Enter the number of calibration
points for each direction (compression and tension or clockwise and counter-clockwise). At least
one point must be selected for each direction. For single-direction sensors such as Omega’s
Series RLC02, only one direction is allowed.
Note:
To achieve the accuracy specification of ±0.1% + sensor, it is recommended to calibrate
the sensor at 5 or more even increments in both the tension and compression directions. For
example, a sensor with capacity of 10 lbF should be calibrated at 2, 4, 6, 8, and 10 lbF loads in
each direction.
2. To escape the
Calibration
menu at any time, press
ESCAPE
. The display appears as follows:
Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/o saving” will return
to the menu without saving changes.
3. After the number of calibration points has been entered, press
ENTER
. The display appears as
follows:
4. Place the force sensor horizontally on a level surface free from vibration, then press
ZERO
. The
indicator will calculate internal offsets, and the display appears as follows:
CALIBRATION
NOT COMPLETE
Cancel
Exit w/o saving
CALIBRATION
Enter # cal points
(1 to 10)
Compression:
5
Tension:
5
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place sensor
horizontally, then
press ZERO.
Содержание DFG-RS5