39
Reprocessing
3. Press the telescope into the notches.
4. Close the instrument tray.
5. Seal the instrument tray in packaging appropriate for
sterilization according to local guidelines and your hospital’s
protocol.
Use of instrument trays
• Do not stack the instrument trays or load them on their side
edges during steam sterilization.
Steam sterilization conditions
Steam sterilization of the product has been validated with
fractionated prevacuum at 132 °C for 3 min.
Other possible steam sterilization parameters are:
Temperature ................................................... >132 °C (269.6 °F)
Holding time ........................................................................ >3 min
NOTICE
Risk of damage to the product
Too high temperatures may damage the product.
• Do not exceed a temperature of 138 °C. Otherwise, damage
may occur.