32
Reprocessing
4. Wipe the eyepiece cover glass and objective cover class with
a cotton swab that has been moistened with 70% ethanol
solution.
Do not wipe with a metal cotton applicator.
5. Thoroughly rinse the product and all gaps at least 10 times
with fresh enzyme-based detergent using a syringe of at least
10 ml.
6. Thoroughly rinse the product with deionized water for at least
30 s using a cleaning pistol or other rinse device.
The cleaning pistol or other rinse device must be suitable for
cleaning medical devices and deliver a minimum pressure of
1 bar (14.5 psi).
7.4 Manual disinfection
NOTICE
Risk of damage to the product
Incompatible disinfectants may considerably damage Olympus
endoscopes and accessories.
•
Use only solutions that are approved and verified (for example
by FDA, DGHM or a comparable institution) and that are
certified by their manufacturers as safe for endoscopic
instrument cleaning and disinfection.
• Select the immersion time and the concentration of the solution
in accordance with the instructions given by the manufacturer
of the disinfectant.
NOTICE
Risk of damage to the product
Disinfectant solutions or tap water may contain various
aggressive compounds that can damage the product.
• To remove residues, rinse the product thoroughly with sterile
water.
• Do not use tap water for rinsing because it might be
chlorinated.
Содержание WA4KL100 ULTRA
Страница 1: ...INSTRUCTIONS FOR USE TELESCOPE ULTRA WA4KL100 WA4KL130 WA4KL145 WA4KL500 WA4KL530 WA4KL545...
Страница 2: ......
Страница 23: ...23 Use Only ignite when the laser emission part is visible on the endoscopic image...
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ...W7510750_06 2018 06 07 en Copyright 2018 Olympus Winter Ibe GmbH...