8.2 Troubleshooting guide
59
MAJ-2306 INSTRUCTION MANUAL
Ch.8
The image (illumination) is
dark or too bright.
The brightness is adjusted improperly.
Set optimum brightness by referring to
“
Adjusting the NBI brightness” on
page 27.
The communication cable is not
connected to the LED light source.
Connect the communication cable by
referring to the instruction manual for
the LED light source.
The light guide cable connected to the
LED light source is disconnected
halfway from the light guide terminal.
Connect the light guide cable
completely by referring to the
instruction manual for the LED light
source.
An extraneous light from a shadowless
lamp, etc. is incident to the observed
region.
Adjust the extraneous light so that it is
not incident to the observed region.
Image is noisy.
The AGC is applied at a high level.
When the image brightness becomes
insufficient in observation of a distant
view, etc., the AGC is applied strongly,
making the image noisy. If this
happens, reducing the observation
distance can increase the image
brightness and decrease noise.
Irregularity description
Possible cause
Solution