4.3 Resetting the current user presets
43
MAJ-2306 INSTRUCTION MANUAL
Ch.4
The selected user setting can be reset by referring to the instruction manual of the surgical
microscope.
After resetting, the settings of NBI become as shown in the following table (see Table 4.12). For details
on other common user setting items, refer to Section 4.4, “Confirming and resetting the system” in the
instruction manual of the surgical microscope.
NOTE
Restoration to the previous setting value is not available after resetting.
Table 4.12
4.3
Resetting the current user presets
Function
Setting
NBI brightness
0
NBI AGC MAX gain
27 dB
NBI red color tone
0
NBI blue color tone
0
NBI chroma
0
NBI color mode
Mode 1
NBI enhancement
Mode 2
NBI Mode 1
Structure enhancement A3
NBI Mode 2
Structure enhancement B3
NBI Mode 3
Edge enhancement E3