6.2 Operation of NBI mode
53
MAJ-2306 INSTRUCTION MANUAL
Ch.6
NOTE
• Yellowish observed regions may look dusky-red during NBI mode.
• The image may sometimes be noisy during NBI mode.
• The contrast cannot be changed during NBI mode.
• For more details on the NBI observation, contact Olympus.
1
By referring to the instruction manuals for the LED light source and surgical
microscope used in combination with this instrument, set the power switch of the LED
light source to ON and then turn ON the surgical microscope.
2
Follow the instruction manual of the ancillary equipment used in combination with this
instrument and turn on the power.
3
Adjust the white balance by referring to the instruction manual for the surgical
microscope.
CAUTION
Be sure to adjust the white balance before use. Otherwise, imaging with correct
color reproduction will be impossible.
NOTE
When the white balance of the WLI mode has been adjusted, the white balance of
the NBI mode is adjusted automatically.
4
Tap the “Pull-down” button for “WLI/NBI” to select “NBI”. Alternatively, press the
custom switch to which the NBI switching function is assigned to select the NBI mode.
Figure 6.1
Pull-down
NBI
Custom button
Example with WLI/NBI on
touch panel
Example with a custom button on
touch panel
Содержание MAJ-2306
Страница 2: ......
Страница 12: ...8 Summary of Equipment Functions MAJ 2306 INSTRUCTION MANUAL...
Страница 14: ...10 1 1 Checking the package contents MAJ 2306 INSTRUCTION MANUAL Ch 1...
Страница 48: ...44 4 3 Resetting the current user presets MAJ 2306 INSTRUCTION MANUAL Ch 4...
Страница 76: ...App 72 EMC information MAJ 2306 INSTRUCTION MANUAL...
Страница 77: ......
Страница 78: ......