
26
Chapter 3 Preparation and Inspection
LF-TP/DP/GP OPERATION MANUAL
3.4
Attaching Accessories to the Endoscope
Attaching the Suction Valve
1.
Insert the Suction Valve’s Main Body into the Suction Valve Housing with the
Suction Lever facing towards the Suction Connector (see Figure 3.9).
2.
While aligning the white line on the Suction Valve Holder with the white dot
on the Suction Valve Housing, depress the Suction Valve Holder until it
stops (see Figure 3.9).
Figure 3.9
3.
Turn the Suction Valve, with the Suction Valve Holder depressed, until it
clicks into position (see Figure 3.10).
Figure 3.10
4.
Confirm that the Suction Lever can be operated smoothly.
Suction Valve
Holder
Содержание LF-TP
Страница 2: ......
Страница 4: ...Contents ii LF TP DP GP OPERATION MANUAL Appendix 49 System Chart 49...
Страница 15: ...Chapter 1 Checking the Package Contents 11 LF TP DP GP OPERATION MANUAL...
Страница 56: ......
Страница 57: ......
Страница 58: ......