FVMPE-RS
3
I Removing the nosepiece
1) Remove the connector attached to the cable of the nosepiece from the connector secured to the left side of
FV30-RFABXL with the binder.
2) With the motorized focusing knob of the controller (U-MCZ), move the stage apart from the objectives
without hiding the nosepiece mounting screw
1
.
3) Fully loosen the nosepiece mounting screw
1
using the Allen screwdriver provided with the system.
4) Gently remove the nosepiece from the front side along the nosepiece mount dovetail.
1
Work related to the nosepiece
1 Removing and mounting the nosepiece
2
Service work
Confirm that the laser is not emitted, and then do the following work.
CAUTION
2-1 Work related to the microscope
} When a nosepiece of a different type from the removed nosepiece is mounted, it is necessary to change
the nosepiece information settings for the touch panel controller. For details, refer to Online Help of the
touch panel controller.
II Mounting the nosepiece
1) With the motorized focusing knob of the controller (U-MCZ), move the stage apart from the objectives
without hiding the nosepiece mounting screw
1
.
2) Fully loosen the nosepiece mounting screw
1
using the Allen screwdriver provided with the system.
3) Gently insert the nosepiece from the front side along the nosepiece mount dovetail, and push the
nosepiece until it stops.
4) Tighten the nosepiece mounting screw
1
.
5) Connect the connector attached to the cable of the nosepiece to the connector secured to the left side of