Changing the halogen bulb
} The designated bulb model is the 12V100WHAL-L (PHILIPS 7724) or
12V50WHAL-L (LIFE JC) halogen bulb.
Fully loosen the clamping screw
➀
at the top of the halogen lamp
housing using the Allen screwdriver provided with the system.
Lift up the halogen lamp housing
2
to remove it.
Tilt the bulb socket
3
by 90° in the direction of the arrow.
While pushing down the bulb clamping lever
4
, hold the halogen
bulb
5
with gloves or a piece of gauze, insert the bulb pins
6
straight and fully into the pin position
7
on the lamp socket.
Then return the lamp clamping lever gently back to the original position
to clamp the bulb.
1
2
3
4
Do not touch the lamp directly.
If fingerprints, etc. are attached, wipe them off completely with
the soft cloth, etc. to prevent the lamp from life reduction or
burst.
CAUTION
Fit the halogen lamp housing from up and tighten the clamping screw
1
by applying downward pressure.
5
1 Changing the lamp
• The bulb, lamp housing and areas near these will be extremely hot during and right after use.
Before replacing the bulb, exit the system and wait until the bulb and the lamp housing are fully
cooled down. Then, replace to the appropriate bulb.
CAUTION
I Changing the halogen bulb
2-2 Work related to the illumination
1
2
3
4
5
6
7
7