•  Main duties of the laser safety officer to manage lasers are as follows.

 

Planning and implementing measures for preventing damages due to laser light 

 

Setting and managing laser controlled area 

 

Managing key and so on which starts laser equipment 

 

Inspecting and maintaining laser equipment, and preserving record of them 

 

Inspecting and maintaining protection tool, and observing if it is being used properly 

 

Conducting labor hygiene education and preserving record of it 

 

Other items necessary for preventing damages due to laser light

•  Service work is work that can affect the performance of the device and such work is described in this 

service manual.  
Only Olympus service personnel or technical personnel who are provided the service training by 
Olympus are allowed to do the service works according to the procedures described in this service 
manual.

•  Doing service work by any other procedures than those given in this manual will result in exposure to 

dangerous laser radiation.

•  During performing service work, do not allow others to enter the periphery of this system except 

service workers.

CAUTION

Содержание FLUOVIEW FVMPE-RS

Страница 1: ...OVIEW FVMPE RS Multi Photon Laser Scanning Microscope Notice Thank you for your purchase of Olympus microscope at this time Retain this manual in an easily accessible place near a system for future re...

Страница 2: ......

Страница 3: ...service manual provided by Olympus The performance has influence on the feature of this system and there is a risk that unintended CLASS 4 laser light is emitted In order to maintain the full perform...

Страница 4: ...trademarks of their respective owners Caution If the system is used in a manner not specified by this manual the safety of the user may be imperiled In addition the system may also be damaged Always...

Страница 5: ...to the nosepiece 3 2 2 Work related to the illumination 7 1 Changing the halogen bulb 7 2 Changing the mercury lamp 9 2 3 Work related to optical unit 11 1 Fluorescence mirror unit 12 2 DM unit for th...

Страница 6: ......

Страница 7: ...piece caps and or nosepiece as this causes irradiation of laser beams that are parallel not diffused This system uses an IR pulse laser which has an element of danger to cause blindness if it enters t...

Страница 8: ...erving record of it Other items necessary for preventing damages due to laser light Service work is work that can affect the performance of the device and such work is described in this service manual...

Страница 9: ...g work CAUTION 2 1 Work related to the microscope When a nosepiece of a different type from the removed nosepiece is mounted it is necessary to change the nosepiece information settings for the touch...

Страница 10: ...4 BX63LF FV30GF...

Страница 11: ...d objective lens LUMPLFLN60XW LUMPLFLN40XW XLPLN25XWMP2 XLPLN25XSVMP2 XLSLPLN25XSVMP2 When using the objective lens marked with use the RMS adapter 4 provided with the piezo nosepiece adapter between...

Страница 12: ...zo nosepiece 5 Connect the piezo nosepiece exclusive controller to FV30 ANALOG with the commercially available BNC cable Connect the control signal input connector CONTROL IN of the piezo nosepiece ex...

Страница 13: ...the lamp socket Then return the lamp clamping lever gently back to the original position to clamp the bulb 1 2 3 4 Do not touch the lamp directly If fingerprints etc are attached wipe them off comple...

Страница 14: ...he lamp housing c a b a b c d Remove the collector lens of the U LH100IR lamp housing a by loosening the three clamping screws b with an Allen wrench width across flats of 2 5 mm 1 While positioning t...

Страница 15: ...g the lamp connector a Hold the black frame and raise the lamp slowly upward If the lamp is dropped the glass will break and you may be injured Pay careful attention when you replace the lamp Do not t...

Страница 16: ...t method 1 30 Press the reset switch for two seconds or more 1 40 or later Press the reset switch and the lamp switch at the same time for five seconds or more 9 c c 9 9 Version 1 30 Version 1 40 or l...

Страница 17: ...t U FDICTIR IX3 FDICT FV30 RFABXL DM unit for SIM scanner FV30 ADM800 FV30 ADMOPT FV30 HSUMP DM unit for External NDD Non Descanned Detector FV30 NDM690 FV30 NDM760 FV30 NDMVCOIR FV30 RFABXL Optical u...

Страница 18: ...disabled and the optical performance of LSM could not be guaranteed Before attaching or detaching the fluorescence mirror unit be sure to engage the shutter in the light path Be sure to exit the FV30...

Страница 19: ...o the space where the fluorescence mirror unit was removed until it touches the end 2 CAUTION 3 4 Hold the both right and left sides of the fluorescent mirror unit with your thumb and the index finger...

Страница 20: ...HSUMP Attaching and removing the cover 1 Remove the cover of FV30 HSUMP by loosening the cover fixing screws four locations of FV30 HSUMP using the Allen screwdriver provided with the system Cover fix...

Страница 21: ...mount dovetail of the DM into the mounting frame of FV30 HSUMP unit until it meets the end of the frame Tighten the screw securing the DM using the Allen screwdriver provided with the system 6 Loosen...

Страница 22: ...RFABXL Attaching and removing the cover 1 Remove the cover of FV30 RFABXL by loosening the cover fixing screws four locations of FV30 RFABXL using the Allen screwdriver provided with the system Cover...

Страница 23: ...the frame Tighten the screw securing the DM using the Allen screwdriver provided with the system DM for external NDD Screw securing the DM If the power of the system is turned ON check that the syste...

Страница 24: ...he laser port FV30 LPADP When FV30 LPADP is placed to laser ports of FV30 HSUMP and FV30 SIMMP the secured cover can be removed with the screw M3 IR port FV30 SIMMP Vis port FV30 SIMMP IR port FV30 HS...

Страница 25: ...less against the 8 hole of FV30 LPADP Tilt 90 0 6 against the surface to which FV30 LPADP is pushed The tolerable power density is the value when the incident position accuracy is satisfied If you ob...

Страница 26: ......

Страница 27: ......

Страница 28: ...4 8F Olympus Tower 446 Bongeunsa ro Gangnam gu Seoul Korea 135 509 48 Woerd Avenue Waltham MA 02453 U S A Distributed by Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan Manufac...

Отзывы: