Rotate the turret by the touch panel controller or the U-MCZ while
checking the display area a of the fluorescence mirror unit No. and
move the fluorescence mirror unit to be replaced to the position on
the front side of the cover.
When rotating the turret, put your fingers away from the turret.
Otherwise, your fingers may be caught.
Hold the sides of the fluorescence mirror unit to be replaced, and
remove it by pulling toward the front side.
Push the fluorescence mirror unit to be used into the space where the
fluorescence mirror unit was removed until it touches the end.
2
CAUTION
3
4
Hold the both right and left sides of the fluorescent mirror unit with
your thumb and the index finger, and shake it gently to make sure that
the fluorescent mirror unit is fixed securely. Without this process, the
fluorescent mirror unit may be attached keeping tilted.
5
After attaching or removing the part, reposition the cover and secure
it by tightening the cover fixing screws using the Allen screwdriver
provided with the system.
6
} Put the hook inside the cover into the groove of the microscope
frame.
} If the power of the system is turned ON, check that the system
starts initializing when the cover is attached.