FVMPE-RS
9
2
Changing the mercury lamp
• Before replacing the lamp, be sure to turn (OFF) the power
switch and unplug the power cord from the outlet, and wait
(more than 40 minutes as a guide) until the lamp replacement
cover and the lamp are cooled down sufficiently. If you replace
the lamp keeping the power switch “ I ” (ON), the lamp may be
damaged.
• Wear protective gloves and protective glasses when replacing
the lamp.
Use a coin or similar to unlock the lamp replacement cover.
Lift the lamp replacement cover to remove the cover.
Unplug the lamp connector
a
.
Hold the black frame and raise the lamp slowly upward.
If the lamp is dropped, the glass will break and you may be
injured. Pay careful attention when you replace the lamp.
} Do not touch the glass area of the lamp (front glass or reflector)
directly with bare hands. The lamp may not be used due to the
attached dirt or fingerprints.
Insert a new lamp to the light source unit along the slide rail
b
.
Connect the lamp connector
a
.
Mount the lamp replacement cover and lock the lamp replacement
cover with a coin or similar.
Set the power switch to “ I ” (ON). The version of the software is
CAUTION
1
2
3
4
CAUTION
5
6
7
8
2
1
3
4
5
6
b
a
7
<< Applicable lamp >>
DC130W SHI Lamp (USHIO INC. SHI-130 OL)
1 Replacement of the lamp