6. Turn on processor and light source. If using master control on the trolley press button in
middle of top shelf front edge, or turn on at video processor and light source.
7. Lamp will ignite automatically or to ignite light lamp – press and hold lamp “Standby
switch” until it beeps”. Lamp turns on and green light next to “ON” illuminates”.
8.
“White Balance Incomplete”
may appear on the screen. To white balance endoscope,
place distal tip of the endoscope into white balance barrel (cap). Press and hold down
the “White Balance” button until you hear a beep and the words “White Balance
complete” appears on the screen.