8. Attach the Injection Tube (MH-946) (Octopus) to the Light Control outlets (Suction,
air/water, air). For those endoscopes with extra channels attach extra injection tubes.
9. If you have a colonoscope or gastroscope with an auxiliary water channel, attach the
corresponding cleaning tube - Auxillary water tube (MAJ-855) to the channel opening
and inject 90ml of enzymatic solution.
10. Place the filter end of injection tube in a bowl of fresh enzymatic solution, and flush each
channel with 90mls of enzymatic solution.