Balloon Water Feeding and Aspiration for the BF-UC180F
Issue
Cause
Solution
Water feeding to the balloon
and/or aspiration from the
balloon does not work.
Balloon water supply channel is dirty.
Reprocess the channel.
Monitor
Issue
Cause
Solution
Why do I have a poor image on
my screen?
Check video cable is correctly plugged in
Check electrical connection
Check white balance
Image on screen, possibly
flickering
Check video cable is correctly plugged in
Check electrical connection
Why do I have lines on the screen
Check video cable is correctly plugged in
Check electrical connection
The image is poor and yet I have
correct colour
Check video cable is correctly plugged in
Check electrical connection
Check white balance
No image on screen
Check video cable is correctly plugged in
Check electrical connection
Check monitor on correct input
Check Power switch on monitor
Check main power source is on
Check that monitor contrast has not been
changed
blown light lamp in light source
Replace light lamp as per manufacturer’s
guidelines
CCD may be damaged
CCD needs replacement as per
manufacturer’s guidelines. Return to
Olympus for repair.
Damaged video cable in the 180 series
Replace video cable
Video cable in the 180 series placed
upside down
Replace video cable in the correct
position