E303 reformat portable
memory
If the portable memory still cannot be
formatted, replace it with a new one.
E311
WHT BAL incomplete E311
Adjust WHT BAL
The white balance has not
been set.
Set the white balance as described in manual
Section 6.5, “White balance adjustment” on
page 224.
Leak Detected in the Endoscope
Issue
Cause
Solution
What do I do if I have detected a
leak in my endoscope?
External Leak can occur in one or more
of the following areas A-rubber,
insertion tube, universal cord, control
body
Call service centre on
1300 132 992
for
further advice.
If leak is in obvious external location:
•
Remove from water
•
Dry area thoroughly
•
‘Patch’ leak with plasticised tape
(“Sleek” or similar)
•
Leak test again to confirm affected
area is patched effectively
•
Continue with reprocessing as usual
•
Package endoscope in yellow
infectious waste bag inside scope
case
•
Transport to service centre after
notifying Olympus
Internal leak can occur in one or more
of the following areas:
Biopsy channel, suction channel,
air/water channel, suction cylinder,
air/water cylinder
Internal Leak
Call service centre on
1300 132 992
for
further advice.
If leak is internal (i.e. in channel):
•
Keep leak tester attached and
running
•
Continue with manual cleaning as
far as possible
•
Brush / flush / Alcohol
•
Package endoscope in yellow
infectious waste bag inside scope
case
•
Transport to service centre after
notifying Olympus
All information are adapted from the following Olympus manuals:
•
CV-180, CV-190
•
CLV-180, CLV-190
•
GIF-180, GIF-190 and GIF-Q190
•
CF-HQ190, CF-H190, CFH180, CFH180AL/I, CFQ180AL/I, and H180AL/I
•
BF-UC180F
•
SIF-Q180
QR 07.2
81
V1.0
November
2019
For
complete
product details see Instructions for Use.