ENF-V3 OPERATION MANUAL
Guidance and manufacturer’s declaration — Electromagnetic
immunity
This model is intended for use by medical personnel in healthcare facility environments and for
use in the electromagnetic environment specified below. The customer or the user of this model
should assure that it is used in such an environment.
Immunity
test
IEC 60601-1-2
(2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic environment —
Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Contact:
±
8 kV
Air:
±
2,
±
4,
±
8,
±
15 kV
Contact:
±
2,
±
4,
±
6 kV
Air:
±
2,
±
4,
±
8 kV
Same as left
Floors should be made of wood, concrete,
or ceramic tile that hardly produces static.
If floors are covered with synthetic material
that tends to produce static, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV
for power supply
lines
±
1 kV
for input/output
lines
±
2 kV
for power supply
lines
±
1 kV
for input/output
lines
Same as left
Mains power quality should be that of a
typical commercial (original condition
feeding the facilities) or healthcare facility
environment.
Surge
IEC 61000-4-5
Differential mode:
±
0.5,
±
1 kV
Common mode:
±
0.5,
±
1,
±
2 kV
for signal input/
output lines:
±
2 kV
Differential
mode:
±
0.5,
±
1 kV
Common mode:
±
0.5,
±
1,
±
2 kV
Same as left
Mains power quality should be that of a
typical commercial or healthcare facility
environment.