32
3.6 Inspection of the endoscopic system
ENF-V3 OPERATION MANUAL
Ch.3
Inspection of the ancillary equipment
Inspection of the endoscopic image
Confirm that the WLI and NBI endoscopic images are normal.
WARNING
Do not stare directly into the distal end of the endoscope while the examination light
is ON. Eye injury may result.
3.6
Inspection of the endoscopic system
Turn ON the video system center, light source, and monitor. Inspect them as
described in their respective instruction manuals.
1
Before inspection, wipe the objective lens with clean lint-free cloths moistened with
70% ethyl or 70% isopropyl alcohol.
2
Observe the palm of your hand in the WLI and
NBI endoscopic images.
Figure 3.13
3
Confirm that light is output from the endoscope’s distal end. (See Figure 3.13)
4
Adjust the brightness level as appropriate.
5
Confirm that the WLI and NBI endoscopic images are free from noise, blur, fog, or
other irregularities.