2.1 Nomenclature and functions
19
ENF-V3 OPERATION MANUAL
Ch.2
No.
Nomenclature
Description
9
Light guide
Connects the endoscope to the light source and transmits light to the distal end of the
endoscope.
10
Venting connector
Attach the ETO cap or leakage tester here.
11
ETO cap
(MB-156)
The ETO cap is equalizing the outer and inner pressure. The cap must be attached prior
to gas sterilization (ethylene oxide gas, STERRAD
®
50/100S/NX
etc.) and aeration
and removed prior to immersion or clinical examination. The cap must also be attached
when the endoscope is transported outside the healthcare facility (shipment, return for
repairs, etc.).
12
Light guide
connector
The light guide connector connects the endoscope to the output socket of the light
source and transmits light from the light source to the distal end of the endoscope.
13
Universal cord
Connects the light guide connector and the control section.
14
Video cable
Connects the light guide connector and the video connector.
15
Video connector
The video connector connects the endoscope to the output socket of the video system
center so that the endoscopic image becomes visible. The endoscope contains a
memory chip that stores information about the endoscope and communicates this
information to the video system center OTV-S190. For more details, refer to the
instruction manual for the OTV-S190.
16
Electrical contacts
Connects the video system center and the endoscope electrically.
17
UP mark
When the video connector is connected to the video system center, this mark is facing
upward.
18
Serial number
The serial number is marked here.